Measuring and logging signals come in handy!
SHS1000X series features full isolation between the two oscilloscope channels, one multimeter channel, power adapter and the USB host/device port. The full isolation makes it ideal for both laboratory and floating signal measurement because it reduces the risk of accidental short circuits. For SHS1000X series, the max voltage input to the analog scope inputs is CATIII 600 Vrms, CATII 1000 Vrms, and the max input for the multimeter is CATIII 600 Vrms, CATII 1000 Vrms. IEC61010-2-030 defines measurement categories that rate the ability of measuring instruments to withstand short-term transient overvoltages outside the operating voltage.
The handheld oscilloscopes feature a sealed IP51 dust and drip-proof housing and have been tested to IEC60529 to guarantee the ruggedness that is needed to survive harsh environments. A rubberized surface with large keys also makes it easy to use in difficult environments.
Main Features
-
Isolation voltage:CATIII 600 Vrms, CATII 1000 Vrms
-
Integrate oscilloscope, serial trigger and decoder, spectrum analysis, data logger and multimeter functions into a convenient and portable design
-
100 MHz or 200 MHz bandwidth models
-
Sampling rate of 1 GSa/s (single-channel), sampling rate of 500 MSa/s (two-channels)
-
Record length up to 12 Mpts
-
Hardware-based true RMS multimeter
-
IP51 certified
Download Here
Specifications
Scope Specifications
Bandwidth
100 MHz, 200 MHz
Sampling Rate (max.)
Two-channel share a single 1 GSa/s ADC
Memory Depth (max.)
6 Mpts/CH (dual-channel mode), 12 Mpts/CH (single channel mode)
Waveform Capture Rate (max.)
100,000 wfm/s (normal mode), 400,000 wfm/s (sequence mode)
Trigger Type
Edge, Slope, Pulse Width, Window, Runt, Interval, Dropout, Pattern, Video
Serial Trigger and Decoder
IIC, SPI, UART, CAN, LIN
Max Input Voltage (scope)
CAT III 600 Vrms, CAT II 1000 Vrms
Between BNC Signal and Protecting Earth;
CAT III 600 Vrms, CAT II 1000 Vrms
Between BNC GND and Protecting Earth;
CAT III 300 Vrms
Between BNC Signal and BNC GND
Probe
PB925
Multimeter Specifications
Maximum Resolution
6000
DC Voltage
60 mV ±1%±15 digit
600 mV-600 V ±1%±5 digit
AC Voltage(45 Hz~400 Hz)
60 mV-750 V ±1.5%±15 digit
DC Current
60 mA-600 mA ±1.5%±5 digit
6 A-10 A ±2%±5 digit
AC Current (45 Hz~400 Hz)
60 mA ±1.5%±5 digit
600 mA ±2%±5 digit
6 A-10 ±3%±5 digit
Capacitance
40 nF ±3%±10 digit
400 nF-400 μF ±4%±5 digit
Resistance
600 Ω-6 MΩ ±1%±5 digit
60 MΩ ±3%±5 digit
Continuity
Continuous beep when resistance < 50 Ω
Other Specifications
Display
5.6-inch TFT-LCD (640x480)
Diode
0~2 V
Power Supply
Rechargeable 6900 mA high-capacity lithium battery
Max Input Voltage (meter)
CAT III 300 Vrms, CAT II 600 Vrms
Weight
Net weight 1.75 kg; Gross weight 3.5 kg
Dimension
276 mm * 168 mm * 68 mm
Data Logger (recorder)
Sample Logger - Max sampling rate: 25 kSa/s, min sampling rate: 1 Sa/s;
Measurement Logger - Max interval: 10 minutes, min interval: 0.1 second.
Max items of logging is 4.
Accessories
Two Passive Probes, Multimeter pen, USB data cable, DC adapter, Power cord
Model | Bandwidth | Channels | Max real time sampling rate | Multimeter | Isolation level | Memory depth |
---|---|---|---|---|---|---|
SHS1202X | 200 MHz | 2 CH | 1 GSa/s | Voltage, current, resistance and other eight common functions | CAT III 600 Vrms, CAT II 1000 Vrms | 12 Mpts |
SHS1102X | 100 MHz | 2 CH | 1 GSa/s | Voltage, current, resistance and other eight common functions | CAT III 600 Vrms, CAT II 1000 Vrms | 12 Mpts |